JIA-928DEFLECTION TEMPERATURE BATH |
Put specimen on two supporting racks of testing bath adding load bar heating with speed of 2℃/min. When center of specimen falling down at specific depth (plastic at 0.26mm, ebonite at 0.25~0.33℃/min), the actual temperature required is deflection temperature. It also tests the cross-breaking strength of specimen at permanent temperature (ebomte at 23℃).
|
SpecSpec:
|
Specimen
|
(L)127X(W)12.7X6~15mm
(L)110X(W)3~4.2X9.8~12.8mm
|
Temp. controller
|
~300℃
Programmable, Regular Temp.
|
Load
|
10, 20, 30, 50, 100, 200, 300, 500g
|
Microgauge
|
0.01~10mm
|
Support
|
Space100±2mm(R)3±0.2mm
|
Conductive heating liquid
|
Glycerine, Silicon oil, lnsulating oil(Self-furnished)
|
Mixer
|
1/35HP
|
Dimension
|
46X32X45㎝ (WXDXH)
|
Weight
|
16 kg
|
|
|
|
|